48T
$299.00
96T
$399.00
Species:
Human
Alternative Name:
Angiotensin-1
Uniport ID:
P01019(34~43aa)
Assay Type:
Competitive
unit:
pg/mL
Standard:
8000pg/mL
Sensitivity:
75 pg/mL
Detection Range:
125-8000pg/mL
Assay Time:
1.5H
Sample Type:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Research:
Endocrinology;Hormone metabolism;
Test Principle:
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Human Ang-Ⅰ. During the reaction, Human Ang-Ⅰ in the sample or standard competes with a fixed amount of Human Ang-Ⅰ on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Human Ang-Ⅰ. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Human Ang-Ⅰ in tested samples can be calculated by comparing the OD of the samples to the standard curve.
| pg/mL | OD | Corrected |
|---|---|---|
| 8000 | 0.226 | |
| 4000 | 0.383 | |
| 2000 | 0.593 | |
| 1000 | 0.857 | |
| 500 | 1.141 | |
| 250 | 1.441 | |
| 125 | 1.775 | |
| 0 | 2.571 |
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
| Intra-assay Precision | Inter-assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean(ng/mL) | 235.8 | 1626.6 | 2459.81 | 160.4 | 1534.08 | 2935.59 |
| Standard deviation | 5.83 | 61.55 | 194.67 | 0.49 | 102.02 | 108.63 |
| C V (%) | 5.15 | 3.23 | 4.39 | 7.07 | 6.68 | 7.53 |
| Sample Type | Range(%) | Average Recovery(%) |
|---|---|---|
| Serum (n=8) | 91-105 | 98 |
| EDTA plasma (n=8) | 81-95 | 88 |
| Cell culture media (n=8) | 85-97 | 91 |
| Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
|---|---|---|---|---|
| 1:2 | Range (%) | 85-94 | 85-102 | 95-102 |
| Average (%) | 85 | 97 | 99 | |
| 1:4 | Range (%) | 87-103 | 85-102 | 80-97 |
| Average (%) | 101 | 102 | 94 | |
| 1:8 | Range (%) | 97-105 | 80-92 | 96-104 |
| Average (%) | 100 | 84 | 104 | |
| 1:16 | Range (%) | 80-90 | 87-98 | 85-93 |
| Average (%) | 90 | 97 | 88 |